How our sensors work


To get the best out of our biosensors, please remember: The sensors are fragile –handle with care and do not touch or knock the sensing tip. Once a sensor has been rehydrated, store in buffer at 2-8 C. Wet storage of sensor following rehydration is guaranteed for 5 days. Following rehydration do not attempt to re-dry the sensor. Keep exposure of sensor in the air to a minimum –preferably no more than 30s at a time as prolonged exposure to air can result in loss of sensitivity.
How to use our sensors
- Storage
- Rehydration procedure
- Cycling
- First calibration and sensor test –in vitro
- Checking interferences
- Calibration -in situ
- Post calibration
- More specific details for each sensor type please see Instruction for Use
ATP
ADO/INO/HYP
ACH/CHO
GLU
LAC
Store in Fridge at 2-8 C before use.
Operation of sensor after use-by date is not guaranteed.
Prepare Buffer A: containing 10 mM NaPi buffer, pH 7.4, 100 mM NaCl, 1 mM MgCl2, 2 mM glycerol. Pour a suitable volume of Buffer A into rehydration pot, and immerse the tip of the sensor into the buffer. Leave for at least 10 min.
this will improve sensor sensitivity by up to 3 fold. Cycle the sensor from –500 mV to + 500 mV and back at a rate of 100 mV/s for 10 cycles. Then polarize to +500 mV for calibration.
Test against the specific analyte Place sensor and Ag/AgCl reference electrode in the bath, connect Ag/AgCl reference electrode and sensor to 2-electrode potentiostat and polarize to +500 mV. Allow sensor current to asymptote –Take zero reading. Expose the sensor to analyte.
Typically such as 10μM serotonin, 10μM dopamine and 100μM ascorbate acid
Regular calibration during an experiment is necessary to monitor for possible loss of sensor sensitivity after contact with tissue. Loss of sensor activity can be partially restored by cycling as described above. Sensors should be calibrated out of contact with tissue –as the tissue impedes free access of exogenous ACH or CHO to the sensor surface. If the tissue is in a flow chamber, the simplest procedure is to withdraw the sensor from the tissue and allow a test solution of ACH or CHO (e.g. 10 µM to flow through).
Note that calibration of the sensor in a physiological buffer may differ from that recorded in Buffer A due to the different ionic composition of the buffers. The sensor should always be calibrated in a saline of the composition use in the experiment with tissue.
Choosing the right sensor
DIY your own sensor holders
To DIY your own sensor holders, you will need a cable (thinner than a 1ml syringe), a 1ml syringe, a gold-plated pin (buy from Sarissa) and some superglue and soldering kit.
First, solder the pin onto one end of the cable and feed it through the syringe.
Superglue both ends to stabilized the cable. And done!
